2.6
Examination
procedure
IMPORTANT
- Prerequisite for the use of the device:
Correct installation, configuration and commissioning of the system
by your authorised custo med dealer
Resting ECG
Make sure that the ECG device is connected to the PC
and the power supply of the device is guaranteed.
Check that your patient is lying comfortably and is not cold.
Start custo diagnostic and click:
Examination, Resting ECG, New Resting ECG see
Software description...
.
Spray the contact spray
1)
on the electrode application points.
Place the electrodes on the patient according to the placement diagram,
See
2.7.1 Electrode application
.
Start the recording.
The patient must keep still during the recording process.
Stress ECG
Make sure that the ECG device is connected to the PC
and the power supply of the device is guaranteed.
:
The battery charge should be > 20%
(charge status indicator blue or green, not purple).
Charge the device if necessary or use the mains!
When using an ergometer, make sure that the patient is in the optimal
seating position (the extended leg should be slightly bent).
When using a treadmill, make sure to follow the instructions
in
2.7.2 Instructions for stress ECG with treadmill
.
Start custo diagnostic and click:
Examination, Stress ECG, New Stress ECG
see
Software description...
.
Spray the contact spray
1)
on the electrode application points.
Place the electrodes on the patient according to the placement diagram,
See
2.7.1 Electrode application
.
Wait a few minutes so that the contact between the skin
and the electrodes can develop optimally.
Apply the blood pressure cuff.
Start the recording.
1) Contact spray warning
Before each examination, make sure to spray contact spray on the electrode
application points. Never spray the contact spray onto the electrodes or into the
electrodes or leads. Do not use water or contact gel!
Recommended contact sprays from:
C+V Pharma Depot GmbH or GE Medical Systems
Only use recommended contact sprays.
Residues from other products can damage the electrodes.