MHW 0012 – DK 1713 | Version 004 – 20/04/2020 | custo med GmbH
2
Resting and Stress ECG
Hardware, description of device for custo cardio 400 BT and custo cardio 400 accu
[2] Resting and Stress ECG Hardware, custo cardio 400 devices |
page 19
2.7
Attaching the device to the patient
2.7.1 Electrode
application
Thoracic wall resting and stress ECG, standard according to Wilson
V1 (C1)
red
4. Intercostal space at the right sternal border
V2 (C2)
yellow
4. Intercostal space at the left sternal border
V3 (C3)
green
on left-hand side of the 5th rib between C2 and C4
V4 (C4)
brown
5th intercostal space on the left midclavicular line
V5 (C5)
black
on the left-hand side of the anterior axillary line, at the same height as C4
V6 (C6)
purple
on the left-hand side of the midaxillary line, at the same height as C4
Extremities resting ECG
R
red
right
arm
L
yellow left
arm
F
green
left
leg
N
black
right leg
Stress ECG (lying or standing): extremities
R
red
on the right below the collarbone
L
yellow
on the left below the collarbone
F
green
on the left above the hip
N
black
on the right above the hip
Stress ECG (sitting position): extremities
R
red
attach to the deltoid muscle on the right
L
yellow
attach to the deltoid muscle on the left
F
green
9. Left rib
N
black
9. Right rib