Cardiac Rehabilitation

 Hardware, description of device for custo guard (custo cardio concept)

Recommended system requirements

Computer  

Intel Core i3-CPU with HD graphics 4400; 4 GB RAM 

 

256 GB SSD or SSHD (for single-position systems 2TB HDD) 

 

1 GBit network connection (not for single-position systems) 

 

Fanless Dual-DVI (or DP) graphics card (for CPET) 

 

Windows 8.1 x64 (PRO version for joining a domain)

Ports  

One USB 2.0 port per USB device (preferably not USB 3.0) 

 

One COM port each for ergometer and treadmills (serial) 

 

At least Version 4.0 if Bluetooth is installed 

 

otherwise can be deactivated in the BIOS

Monitor  

20“ TFT with DVI or DP port 

 

Full HD resolution 

 

Dual-TFT for spiroergometry

Printer  

600 dpi 

 

MonocRFome (colour recommended for spiroergometry) 

 

USB 2.0 port or network connection 

 

PCL-enabled (increases printing speed with the suitable driver)

 

General system requirements

Operating system  

Windows 7 SP1 – with current updates 

 

(32 bit and 64 bit operating system) 

 

Windows 8 (32 bit and 64 bit operating system) 

 

Windows 8.1 (32 bit and 64 bit operating system) 

 

Windows 10 (32 bit and 64 bit operating system) 

 

Windows Server 2003 (32 bit and 64 bit operating system) 

 

Windows Server 2008 (32 bit and 64 bit operating system) 

 

Windows Server 2008 R2 

 

Windows Server 2012 

 

Windows Server 2012 R 

 

older versions are not supported

PC  

The PC hardware should meet the minimum requirements of the  

 

operating system used. Provide additional RAM (1 GB) for  

 

custo diagnostic.  Please ensure that there is sufficient free space  

 

on the hard disk for the custo diagnostic evaluations. 

 

The PC must meet the requirements of the safety standard  

 

DIN EN 60950 for information technology equipment.

File sizes of the evaluations  

Holter:  

approx. 15 MB (max. 60 MB) 

 

ABPM:  

approx. 128 KB (max. 512 KB) 

 

Holter-ABPM:   approx. 20 MB (max. 25 MB) 

 

Resting ECG:  

approx. 200 KB (for an ECG of approx. 10 seconds) 

 

Stress ECG:  

approx. 6 MB (for an ECG of approx. 20 minutes) 

 

CPET:  

refer to stress ECG 

 

Spirometry:  

approx. 50 KB (max. 256 KB) 

 

Rehab:  

approx. 6 MB (for approx. 45 minutes of exercise) 

 

MCT:  

approx. 10 – 15 MB for 24-hour ECG

Hardware & ports  

DVD or CD-ROM drive,  

 

USB port

Summary of Contents for Custo Guard 1/3

Page 1: ...ilitation Part 2 Hardware description of device for custo guard custo cardio concept 1 2 3 4 Safety Hardware Software Hygiene Operating characteristics 1 3 channel ECG Cardiac rehabilitation MHW 0010...

Page 2: ...anufacturer reserves the right to change the information in this Operating Manual without prior notice The current version can be downloaded from our web site www customed de CAUTION This Operating Ma...

Page 3: ...ing position on the ergometer 11 2 5 3 Blood pressure cuff 11 2 5 4 Adjusting the length of the custo belt 3 for 3 channel ECG 12 2 5 5 Attaching custo guard 1 without an electrode belt to the patient...

Page 4: ...ept IP65 Manufacturer custo med GmbH Maria Merian Str 6 85521 Ottobrunn Germany CE mark Observe the Operating Manual Separate collection of electrical and electronic equipment do not dispose with dome...

Page 5: ...ply that transmits the data by radio waves custo guard is used in cardiac rehabilitation for wireless patient monitoring The ra dio transmission allows continuous monitoring of patients independent of...

Page 6: ...uard 1 sets parts list custo guard 1 set custo guard 1 1 channel ECG transmitter custo belt 1 red 1 channel electrode belt custo belt 1 3 extender extension piece for the electrode belt Adapter set fo...

Page 7: ...ard 3 set parts list custo guard 3 set custo guard 3 3 channel ECG transmitter custo belt 3 red 3 channel electrode belt custo belt 1 3 extender extension piece for the electrode belt Adapter set for...

Page 8: ...for an ECG transmitter Power supply unit for custo charger single and custo charger multi custo charger multi set custo charger multi charger for six ECG transmitters Power supply unit for custo charg...

Page 9: ...can occur that an ECG transmitter that is indicated to be completely charged green LED shows the status Charging red LED again if it is removed and then reinserted into the charger Rechargeable batter...

Page 10: ...lectrode belt The skin must be free from grease and personal care products The electrode belt must be set up so that it fits tightly and the contact surfaces make good contact but that the patient is...

Page 11: ...eir arms crossed over their chest for example Before starting the exercise always check the ECG in the exercise station over view in custo diagnostic and perform measures to improve the ECG if necessa...

Page 12: ...The length adjustment slider is located directly next to the point where the material becomes wider on the custo belt The elastic excess length is completely pulled through the slider so that the slid...

Page 13: ...maximum length is not sufficient use a custo belt extender extension piece The custo belt extender is adjusted to the minimum length when the slider is locat ed next to the material loop The maximum...

Page 14: ...the cable adapters Place the ECG transmitter in the case and close the case Attach the neck strap to the case Adjust the neck strap to the desired length Clean the electrode application points the ski...

Page 15: ...to guard ECG trans mitter onto the 4 contact points Attach the neck strap to the hooks on the custo ECG cable guard 3 Adjust the neck strap to the desired length Clean the electrode application points...

Page 16: ...re 20 C 40 C conditions Air humidity 10 95 rH Air pressure 700 1060 hPa Classification Device with internal power supply Class IIa Type BF DIN EN 60601 1 DIN EN 60601 2 47 Holter standard as far as ap...

Page 17: ...ing system Windows 7 SP1 with current updates 32 bit and 64 bit operating system Windows 8 32 bit and 64 bit operating system Windows 8 1 32 bit and 64 bit operating system Windows 10 32 bit and 64 bi...

Page 18: ...e directly con nected to the public low voltage power supply network that also supplies buildings used for domestic purposes Immunity tests Electrostatic discharge ESD according to IEC 61000 4 2 Fast...

Page 19: ...custo guard is used the device should be watched in order to prove the intended functions If unusual performance features are observed it may be necessary to take additional meas ures for example reor...

Page 20: ...ated below For transmitters whose maximum nominal power is not indicated in the above table the recommended protec tive distance d can be determined in meters m using the equation affiliated with the...

Page 21: ...21 09 2017 custo med GmbH 2 2 Cardiac Rehabilitation hardware custo guard custo cardio concept page 21 Cardiac Rehabilitation Hardware description of device for custo guard custo cardio concept 2 8 E...

Page 22: ...1696 Version 001 21 09 2017 custo med GmbH 2 2 Cardiac Rehabilitation hardware custo guard custo cardio concept page 22 Cardiac Rehabilitation Hardware description of device for custo guard custo car...

Page 23: ...custo guard 3 10302 1 custo guard 3 10301 R 1 custo belt 3 red for custo guard 3 10118 1 custo belt 1 3 extender for custo guard 1 3 Accessories Part no Set product designation 10118 custo belt 1 3 e...

Page 24: ...Manufacturer s contact details custo med GmbH Maria Merian Stra e 6 85521 Ottobrunn Germany Phone 49 0 89 710 98 00 Fax 49 0 89 710 98 10 E mail info customed de Internet www customed de...

Reviews: