Safety instructions
Patient safety
Without medical safety devices, e.g. medical protector, the PC and all the non-
medical devices connected to the system (e.g. monitor and printer), must
be set up and used at a distance of at least 1.5 m to the patient unit (see the
orange area in the figure), as leakage currents can occur.
Non-medical devices and the patient may not be touched simultaneously
during the examination.
During routine maintenance work to non-medical devices connected to the
system the patient must not be touched (risk of electric shock).
All unconfirmed reports produced by the system should only be considered as
suggestions. For diagnosis and therapy purposes it is essential that the results
are checked and assessed by a qualified physician.
Important notes when handling custo screen 300
Continuous cuff pressure, e.g. exerted by a kinked cuff tube, can cause patient
injuries. If the cuff pressure is continuous, the patient should open the cuff’s
hook and loop fastener or switch off the recorder.
Make sure that the cuff tube does not become crushed or that the cross-
section of the cuff tube is not reduced.
The cuff must not be applied to wounds, open or newly operated areas.
When conducting blood pressure measurements, the function of additional
medical devices, which are used in the vicinity of the blood pressure cuff on
the patient, may be effected.
Safety distances of devices to the patient
1.5 m
1.5 m
1.5 m
1.5 m
1.5 m
2.5 m
2.5 m
1.5 m