18
ABPM with custo screen 300 and custo diagnostic | GEB 0154 – DK 1055 | Version 002 – 05.06.2013 | custo med GmbH
Hygiene
04
04.1
Cleaning and disinfection
Important notes
Use only cleaning agents and disinfectants approved by custo med.
Unsuitable agents may damage the device.
Follow the manufacturer's instructions
(e.g. regarding dosage and contact times).
The recorder must never be immersed in liquids or cleaned
with too much water.
custo screen 300
The exterior of custo screen 300 should always be aesthetic and clean. If there
is any dirt, gently wipe down the surfaces with a damp cleaning cloth and an
acid-free cleaning agent.
Light spray application of suitable disinfectants is allowed.
Carrying case and belt
Both case and belt can be machine washed with a mild detergent at 40°C, but
must not be dried in a dryer.
Light spray application of suitable disinfectants is allowed.
Blood pressure cuff
After each use, the blood pressure cuff should be cleaned to remove dirt and
sweat. Observe the instructions on the instruction leaflet provided with the
blood pressure cuff.
Cleaning and disinfection:
Sponge cuff with a damp cloth. If necessary,
removebladder and wash cover with soap or detergent disinfectant.
The cuffs have been tested to withstand the following recommended disinfec-
tants: Cidex, Sporicidin, Microzid, Isopropyl-alcohol 70%, Ethanol 70%, Buraton
liquid. After disinfection the cuff cover should be rinsed and allowed to air dry.
Sterilisation:
The cuff has been tested to withstand Ethylene Oxide (EtO)
gas sterilization. Be sure that all safety precautions regarding aeration after EtO
exposure are followed. Never autoclave.