Blood pressure values exceed limit values
Sys: < 70 mmHg > 270 mmHg
Dia: < 40 mmHg > 155 mm Hg
Sys - Dia: < 15 mmHg, HF: < 35/min > 220/min
Measurement is repeated automatically
Pressure discharge rate outside the given limits
Cause: valve is leaky or defective
Customer service required
Disturbed measurement
Cause: too many movement artefacts
Attach the cuff carefully
Keep the arm steady during measurement
Pressure increase is too slow
Cause: cuff is not tight (defective), valve is leaky
Check the cuff (O ring in connector)
Customer service required
Pressure increase is too fast
Cause: cuff tube is kinked, valve system is clogged
Align the cuff tube
Otherwise contact customer service
Discharge period is too long
Cause: cuff tube is kinked, valve is defective
Customer service, if multiple occurrences during a recording
Error when determining diastole
Cause: cuff has been attached incorrectly / insufficient oscillation
(marking on the cuff is not situated on the artery)
Attach the cuff carefully
Keep the arm still during measurement
Error when determining systole
Cause: systole lies above the pump up pressure,
movement artefacts
Attach the cuff carefully
Keep the arm still during measurement
07.2
Error codes and their causes *)
*) A repeat measurement is automatically performed two minutes after a failed measurement.
Customer service is always required for all error codes which are not listed. Please contact your authorised custo med dealer.