40
ABPM with custo screen 300 and custo diagnostic | GEB 0154 – DK 1055 | Version 002 – 05.06.2013 | custo med GmbH
07
Product information
07.5
Technical data and system requirements
Technical data of custo screen 300
Measurement method
Oscillometric measurement procedure
automatic zero balancing
Measurement range
Heart rate
35 – 220 beats/min
Systolic blood pressure
70 – 270 mmHg
Diastolic blood pressure
40 – 155 mmHg
Max. number of measurements 512
Max. time of recording
72 hours
Duration of a measurement
approx. 30 seconds
Measurement intervals
Defined by the examination settings in custo diagnostic
Standard setting: every 15 min during the day, every 30 min during the night
Intervals can be set between 5 and 90 min
Individual adaptation of all phases (day, night, additional) possible
Cuff pressure
max. 300 mmHg
Cuff sizes
Small (Children)
20 – 24 cm
Default (in the scope of delivery)
24 – 32 cm
XL
32 – 40 cm
XXL
38 – 50 cm
Data transfer
Infrared interface custo com IR, with USB port
(IrDA
standard)
Voltage supply
3 Mignon 1.5 Volt, Type AA
3 accumulators, Ni-MH, 1,2 Volt, at least 1500 mAh
Operating conditions
Temperature
+10°C ... +40°C
Air humidity
30 ... 80% rH
Air pressure
700 ... 1060 hPa
Transport and storage
Temperature
-20°C ... +45°C
conditions
Air humidity
30 ... 80% rH
Air pressure
700 ... 1060 hPa
Dimensions Size
approx.
100
* 66 * 26 mm (L * B * H)
Weight
approx. 112 g (with batteries)
Classification
Device with internal power supply
Class
IIa
Type
BF
DIN EN 60601-1, DIN EN 1060-1 to 1060-4, IEC 80601-2-30
Summary of Contents for custo screen 300
Page 2: ......