42
ABPM with custo screen 300 and custo diagnostic | GEB 0154 – DK 1055 | Version 002 – 05.06.2013 | custo med GmbH
Product information
07
Manufacturer’s Declaration – electromagnetic emissions
The custo screen 300 ABPM recorder is designed for operation in the electromagnetic environment
stated below. The customer or user of the custo screen 300 should make sure that it is used in such an
environment.
Manufacturer’s Declaration – electromagnetic immunity
The custo screen 300 ABPM recorder is designed for operation in the electromagnetic environment
stated below. The customer or user of the custo screen 300 should make sure that it is used in such an
environment.
COMMENT: U
T
is the net AC voltage before applying the test levels
07.7
Manufacturer declaration regarding EMC (electromagnetic compatibility)
according to DIN EN 60601-1-2:2007
Emission Measurements
HF emissions according to CISPR11
HF emissions according to CISPR11
Harmonics according to IEC61000-3-2
Voltage fluctuations/flickers according to IEC61000-3-3
Compliance
Group 1
Class B
Not applicable
Not applicable
Immunity Tests
Electrostatic
discharge (ESD)
according to IEC 61000-4-2
Fast transient electric
interference factors/bursts
according to IEC 61000-4-4
Surges
according to IEC 61000-4-5
Voltage drops,
short-time interruptions
and fluctuations
in the supply voltage
according to
IEC 61000-4-11
Magnetic field with
supply frequency
(50/60 Hz)
according to IEC 61000-4-8
IEC 60601 Test Level
± 6 kV contact discharge
± 8 kV air discharge
± 2 kV for net wires
± 1 kV for input and
output wires
± 1 kV push-pull voltage
± 2 kV push-push voltage
< 5 % U
T
for 0.5 period
(> 95 % drop)
40 % U
T
for 5 periods
(60 % drop)
70 % U
T
for 25 periods
(30 % drop)
< 5 % U
T
for 5 s
(> 95 % drop)
3 A/m
Compliance Level
± 6 kV contact discharge
± 8 kV air discharge
Not applicable
Not applicable
Not applicable
3 A/m
Electromagnetic Environment - Guidelines
custo screen 300 uses HF energy only for its internal func-
tion. Its level of HF emission is therefore very low and is
unlikely to be sufficient to interfere with other electronic
devices.
custo screen 300 is suitable for use in all establish-
ments, including domestic establishments and those
directly connected to the public low voltage power
supply network that supplies buildings used for do-
mestic purposes.
Electromagnetic Environment - Guidelines
Floors should be made of wood or concrete or be
equipped with ceramic tiles. If the floor is provided
with synthetic material, the relative air humidity must
be at least 30 %.
The quality of the supply voltage should correspond
to the one of a typical business or clinical environ-
ment.
The quality of the supply voltage should correspond
to the one of a typical business or clinical environ-
ment.
The quality of the supply voltage should correspond
to the one of a typical business or clinical environment.
If the user of the custo screen 300 requires continued
function, even if interruptions in the energy supply
occur, it is recommended to supply the custo screen
300 from an interruption-free power supply.
Magnetic fields with net frequency should correspond
to the typical values, as they can be found in the busi-
ness and clinical environment.