Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13
RPM
Here you determine the RPM limits. When these limits
are exceeded, the LEDs for high or low speed (RPM) will
illuminate.
Select the value to change (Min. or Max.) and confirm with
SELECT.
Using the arrow keys, change the value and save the new
value with SELECT.
S
ETTING
THE
RPM
LIMIT
VALUES
S
ETTING
THE
PULSE
READOUT
Note
•
The limits selected in this menu only apply to the load range
between 6 and 150 watts. At higher loads the RPM limits
automatically adapt to the respective loads:
Load (watts)
Green RPM range (1/min)
6 - 150
54 - 64 (
adjustable
)
151 - 250
58 - 65
251 - 350
68 - 75
351 - 450
78 - 85
451 - 550
88 - 95
551 - 650
98 - 105
651 - 750
108 - 115
751 - 850
118 - 125
851 - 950
> 125
951 - 999
> 130
P
ULSE
D
ISPLAY
The pulse readout on the display can be turned off.
RPM
Min.
0 ... 70
54 / min
Max. 5 ... 130
65 / min
Select
Display
Puls Display
On
Off
Select