60
Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13
E
LECTROMAGNETIC
C
OMPATIBILITY
EN 60601-1-2
Changes or modifi cations to this system not expressly
approved by custo med may cause EMC issues with this or
other equipment.
This system is designed to comply with applicable regula-
tions regarding EMC.
Its compliance with these requirements has been verifi ed.
It needs to be installed and put into service according to
the EMC information stated as follows.
Warning
• RF Interference •
•
Use of portable telephones or other radio frequency (RF)
emitting equipment near the system may cause unexpected or
adverse operation.
Caution
• Equipment Malfunction •
•
The equipment or system should not be used adjacent to,
or stacked with, other equipment. If adjacent or stacked use is
necessary, the equipment or system should be tested to verify
normal operation in the confi guration in which it is being used.
G
UIDANCE
AND
M
ANUFACTURER
’
S
D
ECLARATION
– E
LECTROMAGNETIC
E
MISSIONS
The ec5000 ergometer is intended for use in the electromagnetic environment specifi ed below. It is the responsibility
of the customer or user to ensure that the ec5000 ergometer is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment - Guidance
RF emissions to EN 55011
Group 1
The ec5000 ergometer uses RF energy only for its
internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions to EN 55011
Class B
The ec5000 ergometer is suitable for use in all estab-
lishments, including domestic and those directly con-
nected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions to EN 61000-3-2
Class A
Voltage fl uctuations/fl icker emissions
to EN 61000-3-3
Complies
Summary of Contents for Ergometer custo ec5000
Page 1: ...EXT 0001 DK 1585 Version 001 2015 11 13 EL 20100043500 Operating Manual Ergometer custo ec5000...
Page 4: ...4 Ergometer custo ec5000 EXT 0001 DK 1585 Version 001 2015 11 13...
Page 6: ...6 Ergometer custo ec5000 EXT 0001 DK 1585 Version 001 2015 11 13...
Page 34: ...34 Ergometer custo ec5000 EXT 0001 DK 1585 Version 001 2015 11 13...
Page 54: ...54 Ergometer custo ec5000 EXT 0001 DK 1585 Version 001 2015 11 13...