61
Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13
G
UIDANCE
AND
M
ANUFACTURER
’
S
D
ECLARATION
– E
LECTROMAGNETIC
I
MMUNITY
The ec5000 ergometer is intended for use in the electromagnetic environment specifi ed below. It is the responsibility
of the customer or user to ensure that the ec5000 ergometer is used in such an environment.
Immunity
Test
IEC 60601 Test Level
Compliance
Level
Electromagnetic Environment -
Guidance
Electrostatic discharge (ESD)
to EN 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV
± 8 kV
Floors should be wood, concrete or ceramic tile. If
fl oors are covered with synthetic material, the rela-
tive humidity should be at least 30 %.
Electrical fast transient/burst
to EN 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input and output
lines
± 2 kV passed
Mains power should be that of a typical commer-
cial or hospital environment.
Surge to EN 61000-4-5
± 1 kV diff erential mode
± 2 kV common mode
± 1 kV
N/A
Mains power should be that of a typical commer-
cial or hospital environment.
Voltage dips, short interrup-
tions and voltage variations
on power supply input lines
to EN 61000-4-11
< 5 % UT
(> 95 % dip in UT) for 0.5
cycles
40 % UT
(60 % dip in UT) for 5 cycles
70 % UT
(30 % dip in UT) for 25
cycles
< 5 % UT
(> 95 % dip in UT) for 5 s
< 5 % UT
40 % UT
70 % UT
< 5 % UT
Mains power should be that of a typical commer-
cial or hospital environment. If the user of the
ec5000 ergometer requires continued operation
during power mains interruptions, it is recommend-
ed that the ec5000 ergometer be powered from an
uninterruptible power supply or a battery.
Power frequency (50/60 Hz)
magnetic fi eld
to EN 61000-4-8
3 A/m
passed
Power frequency magnetic fi elds should be at levels
characteristic of a typical location in a typical com-
mercial or hospital environment.
The ec5000 ergometer has no components suscep-
tible to magnetic fi elds.
NOTE:
UT is the a.c. mains voltage prior to application of the test level.
Summary of Contents for Ergometer custo ec5000
Page 1: ...EXT 0001 DK 1585 Version 001 2015 11 13 EL 20100043500 Operating Manual Ergometer custo ec5000...
Page 4: ...4 Ergometer custo ec5000 EXT 0001 DK 1585 Version 001 2015 11 13...
Page 6: ...6 Ergometer custo ec5000 EXT 0001 DK 1585 Version 001 2015 11 13...
Page 34: ...34 Ergometer custo ec5000 EXT 0001 DK 1585 Version 001 2015 11 13...
Page 54: ...54 Ergometer custo ec5000 EXT 0001 DK 1585 Version 001 2015 11 13...