Hardware, description of device for custo flash 500 / 510 / 510V / 501 / 501L
LED display during a normal recorder start
Further LED states, error display
RED
GREEN
RED
GREEN
RED
RED
GREEN
2 x flashing
fast flashing
(2- 3 times / sec.)
8 x flashing
slow flashing
(1 time / 2 sec.)
slow flashing
steady
steady
no LED
custo multiday card was detected.
The custo multiday card configuration is in progress,
the process may take several minutes.
The custo multiday card configuration is complete.
The device is in recording mode.
The rechargeable battery is almost empty.
When starting for the first time:
The custo multiday card is defective, recording not possible.
With very many new starts:
The maximum number of starts has been reached (60).
Recording is only possible after reading the custo multiday card.
Software error, the custo multiday card does not respond
The custo multiday card is full, the maximum recording duration
has been reached or the capacity of the rechargeable battery
is exhausted.