background image

MHW 0007 – DK 1694  |  Version 002 – 15/07/2019  |  custo med GmbH

2

[2] Holter hardware, custo fl ash  500 / 510 / 510V / 501/ 501L  | 

page 9

Holter

 Hardware, description of device for custo flash 500 / 510 / 510V / 501 / 501L

2.5

 

Charging the rechargeable battery, operating the charger

Notes on the rechargeable battery

It is a lithium ion battery without memory effect. The rechargeable battery can be 

charged at any time, irrespective of the discharging times. A fully charged recharge-

able battery lasts for a recording of seven days.

Always use a fresh and fully charged rechargeable battery for a recording.

The rechargeable battery must only be charged at an ambient temperature between 

10 – 45°C. The rechargeable battery must not be charged when exposed to direct 

sunlight or on the radiator.

Charging the rechargeable battery

   Connect the charger to the mains using the power supply unit

   Insert the rechargeable battery into the charger as shown below

   The LED display of the charger shows the state of the rechargeable battery:

   The maximum charging time is approx. 4 hours

   To remove the rechargeable battery, keep pressed the release key 

    on the rear of the charger and pull out the rechargeable battery.

Yellow LED

Green LED

Battery state

without battery

without battery

battery inserted,

charging in progress

battery inserted,

completely charged

Charger state

PSU inserted?

Ready for operation

Summary of Contents for flash 501L

Page 1: ...are Hygiene Operating characteristics for 3 channel holter recordings 24 hours to 7 days MHW 0007 DK 1694 Version 002 15 07 2019 Operating Manual Holter Part 2 Hardware description of device for custo...

Page 2: ...ted into another language without the prior written consent of custo med GmbH The manufacturer reserves the right to change the information in this Operating Manual without prior notice The current ve...

Page 3: ...tery operating the charger 9 2 6 Starting the recorder and applying it to the patient 10 2 7 Patient instructions 12 2 8 Technical data and system requirements 14 2 9 Manufacturer s declaration regard...

Page 4: ...d GmbH Maria Merian Str 6 85521 Ottobrunn Germany CE mark Observe the Operating Manual Separate collection of electrical and electronic equipment do not dispose with domestic waste Recyclable material...

Page 5: ...r use by trained specialist staff or physicians in clinics and medical practices Patients are only allowed to use the recording device after receiving instruction by trained specialist staff Patients...

Page 6: ...p Chest belt for carrying case standard 60 110 cm custo flash 5xx protect custo sensive disposable electrodes Starterkit for custo flash 500 510 510V Charger for custo flash 5xx rechargeable battery R...

Page 7: ...be pressed to mark events during the recording The patient should make a note of the reason for pressing this key in his recording diary e g health problems stress agitation or similar Release key lo...

Page 8: ...cted The custo multiday card configuration is in progress the process may take several minutes The custo multiday card configuration is complete The device is in recording mode The rechargeable batter...

Page 9: ...rechargeable battery must only be charged at an ambient temperature between 10 45 C The rechargeable battery must not be charged when exposed to direct sunlight or on the radiator Charging the rechar...

Page 10: ...sition as shown Insert the programmed custo multiday card into the recorder Make sure that the direction of insertion is correct see the label in the recorder Assemble the rechargeable battery and the...

Page 11: ...e electrodes of other manufacturers can also be used provided that they are approved by the respective manufacturer for this purpose Remove the protective films from the electrodes Apply the top two e...

Page 12: ...lowed The LED display flashes green during recording approximately once every 2 sec onds If this is not the case contact the medical practice To avoid incorrect record ings the LED display should be c...

Page 13: ...Detach the ECG leads from the electrodes do not pull on the cables Remove the neck strap together with the recorder over the head Store the recorder in a safe and dry place Remove the electrodes from...

Page 14: ...on mode rejection at 50 Hz 80 dB Power supply Rechargeable lithium ion battery 3 7 V 1800 mAh Rechargeable battery lifespan approx 500 charging cycles Dimensions approx 95 65 17 mm L W H Patient leads...

Page 15: ...er 2012 Windows Server 2012 R older versions are not supported PC The PC hardware should meet the minimum requirements of the operating system used Provide additional RAM 1 GB for custo diagnostic Ple...

Page 16: ...ion not for single position systems Fanless Dual DVI or DP graphics card for CPET Windows 8 1 x64 PRO version for joining a domain Ports One USB 2 0 port per USB device preferably not USB 3 0 One COM...

Page 17: ...ge fluctuations flickers according to IEC61000 3 3 Compliance Group 1 Class B Not applicable Not applicable Electromagnetic environment guidelines custo flash 500 510 510V 501 501L uses RF energy only...

Page 18: ...0 15 MHz and 80 MHz are 6 765 to 6 795 MHz 13 553 MHz to 13 567 MHz 26 957 MHz to 27 283 MHz and 40 66 MHz to 40 70 MHz The amateur radio bands between 0 15 MHz and 80 MHz are 1 8 MHz to 2 0 MHz 3 5 M...

Page 19: ...s or stacked together with other devices should be avoided as this could result in fault operation If the devices must nonetheless be used as described above this device and the other devices should b...

Page 20: ...1694 Version 002 15 07 2019 custo med GmbH 2 2 Holter hardware custo ash 500 510 510V 501 501L page 20 Holter Hardware description of device for custo flash 500 510 510V 501 501L 2 10 EC Declaration o...

Page 21: ...MHW 0007 DK 1694 Version 002 15 07 2019 custo med GmbH 2 2 Holter hardware custo ash 500 510 510V 501 501L page 21 Holter Hardware description of device for custo flash 500 510 510V 501 501L...

Page 22: ...or custo flash 5xx 55536 2 SD card for custo flash 5xx 256 MB 55543 1 Carrying case for custo flash 5xx 55551 1 Neck strap for custo flash 5xx adjustable 55553 1 Chest belt for carrying case standard...

Page 23: ...to flash 5xx 55551 1 Neck strap for custo flash 5xx adjustable 55553 1 Chest belt for carrying case standard 60 110 cm 55541 7 custo flash 5xx protect 40007 1 PCK30 custo sensive disposable electrodes...

Page 24: ...Manufacturer s contact details custo med GmbH Maria Merian Stra e 6 85521 Ottobrunn Germany Phone 49 0 89 710 98 00 Fax 49 0 89 710 98 10 E mail info customed de Internet www customed de...

Reviews: