Holter ABPM
Hardware, description of device for custo screen 400
MHW 0004 – DK 1663 | Version 001 – 12/01/2018 | custo med GmbH
[2] Holter ABPM Hardware, custo screen 400 |
page 10
Error when reading or writing the time
Internal electric double-layer capacitor (energy store for the time) discharged
Insert batteries, switch device on
Thresholds are fallen below or exceeded
Sys: < 70 mmHg > 270 mmHg, Dia: < 40 mmHg > 155 mmHg
Sys - Dia: < 15 mmHg
HR: < 35/min > 220/min
Measurement is repeated automatically
Pressure discharge rate outside the given limits
Valve is leaking or defective
Customer service
Disturbed measurement
Too many movement artefacts
Cuff slipped or not tight enough
Attach the cuff carefully and keep the arm steady during the measurement
Battery voltage too low
Insert new batteries or freshly charged rechargeable batteries
into the recorder
Pressure sensors transmit different values
Cuff tube kinked
BNC port at the recorder or cuff tube dirty
One of the pressure sensors is faulty
Remove cuff from device, connect it again and repeat procedure
If the error persists, call customer service
Pressure increase is too slow
Cuff is not connected or cuff/valve is leaking/defective
Connect cuff if necessary
Check cuff (sealing ring in connection present/OK?)
If the error persists, call customer service
2.4.3
Error codes and their causes