2
Holter ABPM
Hardware, description of device for custo screen 400
MHW 0004 – DK 1663 | Version 001 – 12/01/2018 | custo med GmbH
[2] Holter ABPM Hardware, custo screen 400 |
page 11
Pressure increase is too fast
Cuff tube is kinked, valve system is clogged
Align the cuff tube
If the error persists, call customer service
Discharge period is too long
Cuff tube is kinked, valve is defective
Customer service if multiple occurrences during a recording
Error when determining diastole
Cuff attached incorrectly, marking is not situated on the artery,
movement
artefacts
Attach the cuff carefully and keep the arm steady during the measurement
Error when determining systole
Cuff attached incorrectly, marking is not situated on the artery,
movement
artefacts
Attach the cuff carefully and keep the arm steady during the measurement
In the event of an erroneous measurement, the measurement is repeated after two
minutes. If errors occur that are not listed here, switch the device off and back on
again. Repeat the desired step. If the error persists, contact your authorised custo
med dealer.