2
Holter ABPM
Hardware, description of device for custo screen 400
MHW 0004 – DK 1663 | Version 001 – 12/01/2018 | custo med GmbH
[2] Holter ABPM Hardware, custo screen 400 |
page 16
Adjustment from maximum length to medium length
Hold the slider with your right hand, grasp the lower layer of material to the left of
the slider and pull it through the slider up to the wide point.
Adjustment from medium length to minimum length
Hold the slider with your left hand and pull the elastic excess length completely to
the right through the slider until the slider is back between the black plastic hook
and the wide material point.
Extension with the custo belt extender
If the maximum length is not sufficient, use a custo belt extender (extension piece).
The custo belt extender is adjusted to the minimum length when the slider is located
next to the material loop. The maximum length is reached when the slider is next
to the black plastic hook. There is no limit on the number of custo belt extenders
that can be used.
Movable back electrode
For petite patients the back electrode is fitted towards the front electrode (see Fig-
ure) so that the back electrode does not rest on the spine. Open the Velcro fastener
and turn the electrode module to the desired position. Close the Velcro fastener
again. The electrodes must always be in full contact with the patient's skin.