Technical data and system requirements
Technical data of the custo screen 400
Measurement method
Oscillometric measurement procedure
Automatic zero balancing
Measurement range
Heart rate
35 – 220 beats/min
Systolic blood pressure
70 – 270 mmHg
Diastolic blood pressure
40 – 155 mmHg
Measurement accuracy
Systole
Diastole
Deviation from the average value -0.5 mmHg
-0.1 mmHg
Standard deviation
4.5 mmHg
3.3 mmHg
Meets the requirements of the ESH International Protocol 2010
Max. number of measurements
500
Max. recording time
72 hours for ABPM recordings
24 hours for Holter ABPM recordings
Duration of one measurement
approx. 30 seconds
Measuring intervals
Can be set to 5 to 90 minutes in the software
Standard: every 15 minutes during the day,
every 30 minutes at night
Cuff pressure
max. 300 mmHg
Cuff sizes
Cuff with retainer standard
24 – 32 cm
Cuff with retainer x-large
32 – 40 cm
Cuff with retainer xx-large
38 – 50 cm
Bandage cuff small, child
20 – 24 cm
Bandage cuff standard
24 – 32 cm
Bandage cuff xx-large
38 – 50 cm
Data transmission
Infrared interface with USB port (IrDA standard)
mini SD card for Holter ABPM recordings
Power supply
3 Mignon 1.5 V, type AA
3 rechargeable batteries, Ni-MH, 1.2 V, min. 1500 mAh
Operating conditions
Temperature
+10°C ... +45°C
Air humidity
10 ... 95% rH
Air pressure
700 ... 1060 hPa
Transport and storage
Temperature
-20°C ... +45°C
conditions
Air humidity
10 ... 95% rH
Air pressure
700 ... 1060 hPa
Dimensions
Size approx. 100 * 66 * 26 mm (L * W * H)
Weight approx. 159 g (without batteries)
Classification
Device with internal power supply
Class
IIa
Type
BF
Underlying standards
DIN EN 60601-1, DIN EN 60601-1-2,
IEC 80601-2-30, DIN EN 60601-2-47