Main Electrical Panel
➢
To lubricate the Z-Stage leadscrew
1. Turn off all electrical power to the system. For more information, refer to the section
2. Open the top cover of the safety enclosure. Disconnect both cables from the sensor.
3. Remove the sensor by loosening the three captive mounting bolts.
4. Slide the sensor off of the two mounting pins.
5. Using a cotton swab stick, apply a small amount of grease to the leadscrew through a hole in
the center of the sensor mounting plate as shown in Figure 94.
Figure 94.
Z-axis Leadscrew
6. Move the Z-stage up and down manually by turning the knob underneath the Z-stage.
7. Slide the sensor onto the bottom pin of the mounting plate.
8. Pivot the sensor to align with the top pin of the mounting plate.
9. Secure the sensor to the mounting plate by tightening the three captive screws.
10. Reconnect the sensor cables on the back of the sensor and tighten the attaching screws.
Warning:
Make sure the system power is turned off before working inside the hood. Gantry
motion can pinch or sever objects in its path.
Top mounting pin
Bottom
mounting pin
Z Stage
mounting
plate
Captive screw
Captive screw
Captive screw
Leadscrew
access hole