Shenzhen Zhiyong Electronics Co., Ltd.
8
1)
EUT power port:
power input port for EUT 1)
Power input port:
power supply input port
2)
Ground port:
the size of the bolt is 8mm, grounding 2)
Ground port:
the size of the bolt is 8mm, grounding
3)
Output port:
interference signal output port
with Type N female port
The bottom metal panel of EM5040C is the reference ground, users can connect it on the ground of
lab with screw.
5.
Construction of EMC test platform
Please read the instruction book carefully, learn the safety knowledge and measure
according to correct step.
The product is designed according to national standard, so theoretically there will be
large leakage of current to ground. A bad grounding can possibly cause deadly electric
shock. As a result, the company requires the users:
The device must be well grounded (There’re grounding point on the side and rear of
the device)
Isolation Transformer must be installed as double protection.
Our company’s EM5060 isolation transformer can fulfill the requirement for most
tests.
Do not open the case and connect lines during operation. Do not use in damp or
explosive environment. Please keep the device surface dry and clean before using it.
Please make sure the product is working in rating voltage and current range.
If there’s anything wrong with the product, please contact us immediately. Please do
not open the shell and try to repair it to avoid unwanted accident.
CYBERTEK