#2: HIGH CALIBRATION
1. Fill the tank to its full capacity.
2. Wait 5-10 minutes to allow the liquid in the tank to settle.
3. Enter Calibration Mode again on Level-Eye
®
:
a. Press the “L” and “R” buttons together for 5 seconds
until the word “CAL” is displayed on the LCD, then
release.
4. Perform High Calibration:
a. As “CAL” is still displayed, hold down the “R” button
for 5 seconds until “Hi” is displayed, then release.
b. Wait for the word “CHi” to display on the LCD screen,
confirming the high calibration is complete.
5. Level-Eye
®
should now display “100%.” Congratulations!
You have a happy tank.
NOTE: If Level-Eye is ever reinstalled and used on another tank,
or with a different liquefied gas, it MUST be re-calibrated to that tank and/
or that specific gas to operate correctly.
#1: LOW CALIBRATION
1. Add 5 lb. (2.25 kg) of liquid to the empty tank.
2. Wait 5-10 minutes to allow the wand to chill.
3. Enter Calibration Mode on Level-Eye
®
:
a. Hold the “L” and “R” buttons
together for 5 seconds until the
word “CAL” is displayed on the
LCD, then release.
4. Perform Low Calibration:
a. As “CAL” is still displayed,
hold down the “L” button for 5
seconds until “Lo” is displayed,
then release.
b. Wait for the word “CLo” to
display on the LCD screen,
confirming the low calibration is
complete.
LEVEL-EYE
®
USER MANUAL
© 2020-09 Cyl-Tec, Inc. All rights reserved.
Level-Eye
®
Troubleshooting
Problem
Cause
Action
LCD does not display
anything
Replace the battery
1. Remove back cover of Level-Eye to access battery.
2. Replace old battery with new one.
3. Replace back cover. Hand-tighten firmly to fully engage waterproof seal.
LCD only displays
“000%” or “100%”
Incorrect calibration A new calibration sequence (both low and high) needs to be performed.
LCD continues to display
“100%” after use
Vessel may have been
overfilled
Level-Eye will not read below 100% until the liquid level dips below the
calibration point. Monitor gauge regularly according to usage for any change.
LCD displays an “L” next
to a number
Normal reading
“L” means “Level” and the number indicates what percentage of liquid gas is
left in the tank.
“No Symbol”
is
displaying
Head is not properly
communicating with
the wand
NOTE: DO NOT perform a calibration while the No Symbol is present!
Check the brass jack connectivity between the head and the wand:
1. Loosen the wing nut at base of head and remove from wand.
2. Check brass connection jacks on both parts for debris or damage.
3. If connection is clear, wait 30 seconds and re-attach the head to the wand.
4. If “No Symbol” is still present, please contact Cyl-Tec customer service.
LCD is displaying
inaccurate readings
Water damage or
moisture
Check for water damage or moisture in the LCD screen or Level-Eye housing.
If damage is severe, contact Cyl-Tec customer service - (888) 429-5832.
Level-Eye
®
Calibration
Level-Eye
®
calibration is a two-step process of setting the liquid level range from an empty
(low) level to a full (high) level for your specific tank size and model.
IMPORTANT! Calibration must be performed DURING the filling process to calibrate correctly.