4.1
Site requirements
Introduction
This section describes the specifications to which the installation site must comply.
Space requirements
Prepare a clean working area on a stable laboratory bench. The bench must comply
with the specifications in the following table.
Parameter
Specification
Minimum bench area for oper-
ating the ImageQuant 800
70 × 60 cm
Free space required around
the imager
At least 40 cm free space in front of the instru-
ment
10 cm free space on the left, rear, and top sides of
the instrument.
30 cm free space on the right side of the instru-
ment.
Load capacity
981 N (100 kg) or higher
Inclination of bench surface
Horizontal ± 2°
WARNING
Access to power cord with plug. Do not block access to the
power cord. The power cord with plug must always be easy to
disconnect.
NOTICE
This equipment is not intended for use in residential environments
and might not provide adequate protection to radio reception in
such environments.
4 Installation
4.1 Site requirements
48
Amersham ImageQuant 800 Operating Instructions 29372604 AC
Summary of Contents for AKTA go
Page 129: ...Pageintentionallyleftblank...