Problem
Possible cause/corrective action
Too high
outlet pres-
sure
•
Check that connected outlet is actually used.
•
Outlet containers are placed too high relative to the pump. Make
sure that the outlet containers are placed as close to
Chromaflow Packing Station as possible and on the same floor
of the building.
•
Tubing to outlet container causes pressure or flow loss. Reasons
may be too long tubing, to small internal diameter, tube may
have a narrow section or is partly plugged. Check also for non-
functioning valve(s).
•
Malfunctioning valve at container battery limit.
Valves
This section outlines troubleshooting for valve related issues.
Problem
Possible cause/corrective action
Chromaflow Packing
Station does not func-
tion at start
•
Valves might not be working. Check that
compressed air supply is sufficient.
•
Reduce the pump setting and press the
START/
RESET button.
Flow not as desired
Check the status of inlet and outlet valves and also that
the flow path is open.
Compressed air
This section outlines troubleshooting for compressed air related issues.
Problem
Possible cause/corrective action
Compressed air failure
during a run
Insufficient air pressure caused by, for instance,
malfunctioning air supply equipment or a leaking air
hose. Rectify the air supply problem.