4.2 How to install the holder
Gathering materials
Gather the following materials to install your Cytiva cassette holder into your cross-flow system:
•
A cross-flow system including a feed reservoir, feed pump, a minimum of two glycerin-filled
pressure gauges, and three flow control valves
•
System piping and fittings to connect the feed pump, permeate collection vessel, and feed
reservoir to the holder. The system piping must end in ½-inch sanitary clamps to connect to the
Kvick Lab holder.
•
Three, fractional sanitary clamps and ½-inch gaskets
Making the connections
Place the Kvick Lab holder on a sturdy lab bench near your cross-flow system. Stabilize and level the
holder using the adjustable feet. Keep piping runs short to minimize holdup volume. Connect the
holder’s feed, retentate, and permeate ports to your system’s piping using the three gaskets and
sanitary clamps.
4 Connecting the Holder to Your System
4.2 How to install the holder
20
Learning to Use the Kvick Lab Cassette Holder User Manual 18117269 AI