Industrial Flow Cell 8 mm
Parameter
Specification
Recommended maximum flow rate
*
Maximum pressure
2 MPa (20 bar, 290 psi up to 40°C
Backpressure
Max. 0.1 bar at 1800 L/h with water at
25°C
Liquid temperature range
4°C to 80°C (at max 0.1 MPa)
Optical path length
1, 2 and 5 mm
Cell volume
2.5 mL
Wetted materials
PEEK (polyetheretherketone)
Titanium (palladium alloy)
Quartz (synthetic fused silica)
PFR-91 or EPDM
pH stability range
1-13, 13-14 (< 1 days exposure)
Chemical resistance
The wetted parts are resistant to
organic solvents and salt buffers
commonly used in chromatography of
biomolecules, except 100% ethyl
acetate, 100% hexane, and 100% tetra-
hydrofuran (THF) or 15% THF in acetoni-
trile
Tubing connections
TC25 connector
Industrial Flow Cell 1 inch
Parameter
Specification
Recommended maximum flow rate
*
Maximum pressure
1 MPa (10 bar, 145 psi) up to 40°C
Backpressure
Max. 0.4 bar at 600 L/h with water at
25°C.
Liquid temperature range
4°C to 80°C (at max 0.1 MPa)
Optical path length
1, 2 and 5 mm
Cell volume
12 mL
8 Reference information
8.1 Specifications
76
Monitor UVis-920 Operating Instructions 29055049 AF
Summary of Contents for Monitor UVis-920
Page 2: ...Pageintentionallyleftblank ...