48
Step 1:
Clean the tip with soap and water, ensuring
the mirror is clean and free of dirt, smudges or any
residue. Avoid using abrasive materials on cloth as this
will scratch the mirror.
Step by step procedure
Step 2:
After drying the outside of the tip, carefully dry
the inside and mirror with lint-free wipes, ensuring you
don't scratch the surface. The mirror must be free of
visible debris or water spots.
Step 3:
Insert and seal the tip in a sterilization package
with the tip facing down so that water droplets do not
accumulate on the mirror. Make sure the seal is airtight.
Each tip must be individually packaged.
6.3. CLEANING AND STERILIZATION OF THE TIP
The included tips must be autoclaved before use as they are not pre-sterilized.
The scanner tip is autoclavable up to 250 times in a steam autoclave.
Make sure the mirror surface does not show any residue, smudges, scratches,
or any damage as this would affect the performance of the device.