Radiographic image scanner EAGLE.PS
27
EQUIPMENT OPERATION
Perform Measurements
The application of measurements and calibrations in the image is available:
Straight:
This feature makes it possible to measure straight in any direction in the image. The unit of
measure is 'mm' (millimeter), with precision of two decimal places.
It is also possible to draw several consecutive lines, here called "Multiple Straight lines", in any direction in
the image, where the system displays the total sum of the drawn lines;
Angle:
This feature makes it possible to perform angle measurements between two consecutive lines.
The unit of measurement shall be 'degrees', to the accuracy of two decimal places;
Calibrate Measurement:
This feature enables to calibrate the image. The system will display a pop-up as
shown In Figure 20, allowing the operator to calibrate the actual measure of the image. The value displayed
in the measure in pixels, refers to the amount of pixels of the image, and therefore cannot have its value
changed.
To perform measurements and calibration, the user must select them through their icons in the Toolbar:
Figure 17 - Toolbar: Measure Picture Image
Figure 18 - Toolbar: Available Measurements
Figure 19 - Toolbar: Measurement Calibration
Figure 20 - Measurement Calibration Pop-up