EQUIPMENT OPERATION
User *:
Represents the name or part of the name of the user logged into the system at the time the
inclusion action is originated in the database
Remarks:
The system will apply the mask "dd / mm / yyyy" to the 'Date *' field. It will also validate it.
The system will apply the mask "hh: mm: ss" to the 'Time *' field
Erase Phosphorous plate
This feature allows the operator to format the phosphorous plate. To perform such procedure, the software
will establish communication with EAGLE.PS.
The user is allowed to erase the phosphorous plate at any time. It should be noted that this operation is not
reversible, i.e., once the phosphorous plate is erased it will not be able to recover an image a new x-ray
should be performed.
Phosphorous plate erasure is available through the following path: Menu Erase plate Home Screen– Other
Options – Erase plate
Remarks:
If the user does not insert the phosphorous plate through the input plate in a maximum period of 120
seconds, the software will display a message that the phosphorous plate has not been inserted.
Software Configuration
These are the features that enable the operator to perform the configurations on the software to log in,
select the system language, configure the Diagnosis header and footer and configure the quality of the
image of the EAGLE.PS equipment.
Software configuration is available through the following path:
Configure Menu – SW Configure
Toolbar – System setup
Login Settings
This is the Login Setup screen. Enables the operator to log in
Figure 65 - Software Settings: Login Option