Radiographic image scanner EAGLE.PS
89
UNFORESEEN- TROUBLESHOOTING
Message
Error or warning grounds
Solution
−
Please enter a valid email
address for contact.
−
The system typically displays
this Message when it fills in the
email field with an invalid email.
−
Fill in the e-mail field with a valid
e-mail.
−
The image size exceeds 500kb.
Should you have any questions,
please contact the System
Administrator.
−
The system usually displays this
Message when the operator
requests to load an image that
exceeds the size of 500Kb.
−
Operator to upload an image
with size less than 500Kb.
Retrieve Image
Message
Error or warning grounds
Solution
−
There are unfilled mandatory
fields.
−
The system usually displays this
Message when the required
fields of the screen in question
are not filled in.
−
All required fields must be filled
in.
−
Current User cannot be an X-
Ray requestor since it is not
classified as a Dentist.
−
The system typically displays
this Message when the operator
attempts to select a user that
does not have the 'Dentist'
profile as the Requestor.
−
The operator must select a user
having "Dentist” profile
−
Initializing equipment, please
wait.
−
The system typically displays
this Message when the operator
tries to retrieve an image when
the machine is initializing.
−
Wait for the equipment to start
up, i.e. when the blue LED is lit
steadily.
−
There was an error accessing
the database. If the problem
persists contact the System
Administrator.
−
The system typically displays
this Message when there is no
connection with the data base.
−
Reestablish communication with
the database.
−
All reported data will be lost. Do
you want to continue?
−
The system typically displays
this Message when the retrieval
procedure has not been
completed and the operator
clicks on any of the system
menu items.
−
Finish the entire image
acquisition process before
accessing other menu item.
−
Communication failure with the
equipment
−
The system usually displays this
Message when a
communication problem occurs
with the equipment
−
Remove the cable and
reconnect it again.