CORRECTIVE AND PREVENTIVE MAINTENANCE AND PRESERVATION
Additional procedures for reuse
The equipment can be reused in undetermined, i.e. unlimited, quantities, only needing to be cleaned
and disinfected.
Cleaning and Disinfection
These processes should be used in the handpieces.
Uncouple the piece from the INTRA fit terminal.
By using a clean, soft cloth, clean the outside by employing alcohol at 70% or neutral soap.
Do not use microabrasive material or steel wool when cleaning, or employ organic solvents
ordetergents which contain solvents such as ether, stain remover, etc.
Spray lubrication
To carry out the lubrication of your equipment, we recommend the use of the Lubs product.
Lubrication must be carried out by the posterior hole of micromotor as illustrated in fig. (a). Firmly
fit the oil applicator nozzle into the hole indicated in fig. (a) and apply the spray for 1 or 2 seconds
(fig. b) by holding the vial in vertical position.
Perform these operations (spray lubrication) with the regulating ring positioned clockwise and
counterclockwise.
Before of use, activate a handpiece during some seconds to eliminate the excess of oil. Cover the
piece with absorbent paper to remove the excess of oil.
Warning
Never submerge the instrument in a disinfection bath.
Never use product peracetic acid for cleaning, since it is corrosive.
The lubricant doesn’t follow the product.
Attention:
The non-use of the lubricant will result
In the loss of product warranty.
Terminal type
BORDEN
Channel To
Lubrication
(Entry of air)
Channel To
Lubrication
(Entry of air)
Terminal type
MIDWEST
fig. a
fig. b
NOTE: Handpieces no submitted to process
of sterilization should be lubricated at
least twice a day.
control ring