15
# Function mode list #
Name
Function
Setting
출고시
기준설정값
F-01 Decimal point
0, 1, 2, 3
1
F-02 Division
1, 2, 5, 10, 20, 50
1
F-03 Display filter
1, 2, 5, 10, 20, 50, 100
20
F-04 Hold mode
Sample hold, Peak hold, Absolute peak hold
Peak hold
F-05 Comparison mode
Decision, High limit, Low limit
Decision
F-06 Hysteresis
0 ~ 99
0
F-07 DAC zero
V_05V, V_10V, V02A, V_42A,
b_05V, b_10V, b_02A, b_42A
V_10V
F-08 DAC capacity
-19999 ~ +99999
10000
F-09 ID Number
0 ~ 32
0
F-10 Baud rate
2400, 4800, 9600, 19200, 38400, 57600
9600
F-11 Auto zero tracking
0 ~ 99
0
F-12 Auto zero tracking time 0.0 ~ 5.0 sec
0.0
F-13 Force unit
Kg(kg/
㎠
), N,
ℓ
b, bar, MPa
Kg
F-14 Key disabling
Zero key, Lo key, Hi key, Hold key
0000
F-50 Offset (Base)
-19999 ~ +10000
0
F-51 Direction
0, 1, 2
0
※
F-07 (OPTION, Analog out)
Display data
Setting
V_05V
0 ~ ±5V Output
V_10V
0 ~ ±10V Output
V_02A
0 ~ 20mA Output
V_42A
4 ~ 20mA Output
b_05V
2.5V ~ 5V Output 2.5V ~ 5V)
b_10V
5V ~ 10V Output (0 ~ 5V ~ 10V)
b_02A
10mA ~ 20mA Output (0mA ~ 10mA ~ 20mA)
b_42A
12mA ~ 20mA Output (4mA ~ 12mA ~ 20mA)