18
F-10. Baud rate & Print (Communication Speed Setup)
(Standard setup value: 9.60)
Display data
Setting
Stream mode
Command mode
2.40
2400 bps
O
O
4.80
4800 bps
O
O
9.60
9600 bps
O
O
19.20
19200 bps
O
X
PRINT DATA OUT (PT-100)
-
-
F-11. Auto zero tracking (Auto zero operation range setup)
(Standard setup value: 00)
Display data
Setting
00
00
: Auto zero is not used
~
99
01 ~99 : Set up the operation range of auto zero
F-12. Auto zero tracking time (Auto Zero tracking time setup)
(Standard setup value: 0.0)
Display data
Setting
00
0.0
: Auto zero is not used
~
5.0
0.1 ~5.0 : set up the auto zero operation time (0.1 ~ 5.0 sec)
F-13. Force unit (Conversion unit set up)
(Standard setup value: 0)
Display data
Setting
0
kg.f
kg/
㎠
1
N
kgf × 9.8
-
-
2
ℓ
b
kgf × 2.2
-
-
3
-
-
Bar
kg/
㎠
× 0.98
4
-
-
MPa
kg/
㎠
× 0.098
Note) During calibration, set up the unit you want after calibrating it into Kg.