8-3. Digital calibration (Calibration by sensor output value)
At the time of purchasing sensor, the rated capacity (R.C) and rated output (R.O)
declared on the calibration sheet can be used for the calibration for easier calibration.
Measuring Mode
While
key is being
pressed, press
key.
Displays Function Mode
Displays Sensor Output Calibration State
Displays Rated Output value Input State
Displays the rated capacity value Input State
Input the rated capacity value by using
and
key. And then press
key.
Calibration Completed.
Input the rated output value by using
and
key. And then, press
key.