5-5. Actual load calibration
This is a calibration method by adding actual load on the sensor. Standard weight is needed.
Display Shunt CAL value (please write to S.CAL)
Cal. Date R
14
Measuring mode
While
key is being Press, press
key
Display Function mode
Display Sensor output value calibration mode
Display Actual load calibration mode
Display actual load mode
Input the rated output value by using
and
key. And
then. Press
key
Calibration end
Display zero mode. press
key
without any load on the
sensor
About 1 sec later
After adding the actual load (standard weight) on the
sensor, press
ke
y.
About 1 sec later
Display R.O. value input mode
Input the rated output value by using
and
key. And then.
Press
key (Input 3.0000 if do not know the R.O. value)
.O. value calibration constant vlaue