Immunity to
Conducted
Disturbances
Induced by
RF fields
EN 61000-4-6:2014
IEC 61000-4-
6:2013
Operating
Stand by
AC Mains
AC 120V,60Hz
AC 220V,60Hz
AC 230V,50Hz
3 V
0.15-80 MHz
6 V in ISM bands
Between 0.15 MHz
and 80 MHz
80% AM at 1 kHz
Power Frequency
Magnetic Field
Immunity
EN 61000-4-8:2010
IEC 61000-4-
8:2009
Operating
Stand by
Enclosure
AC 120V,60Hz
AC 220V,60Hz
AC 230V,50Hz
30 A/m
50 Hz & 60 Hz
Voltage dips
EN 61000-4-
11:2004
IEC 61000-4-
11:2004
Operating
Stand by
AC Mains
AC 100V,50HZ
AC 100V,60HZ
AC 127V,50HZ
AC 127V,60HZ
AC 220V,60HZ
AC 230V,50HZ
0 % U
T
: 0.5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % U
T
; 1 cycle
and
70 % U
T
; 25/30
cycles
Single phase: at 0º
Voltage
interruptions
EN 61000-4-
11:2004
IEC 61000-4-
11:2004
Operating
Stand by
AC Mains
AC 100V,50HZ
AC 100V,60HZ
AC 127V,50HZ
AC 127V,60HZ
AC 220V,60HZ
AC 230V,50HZ
0 % U
T
; 250/300
cycle