Error codes and instructions to follow
1) Code 10
(Cuff connection error)
- During power on the controller performs an automatic one cycle calibration to
detect the cuff type and proper garment fit. If the controller fails to detect the cuff,
it shows Error code 10 with an alarm sound.
*Checking point
①
Check the connection of the controller to the
hoses (tubing).
②
Check that each cuff is wrapped properly on the
patient’s leg.
(3) Error Description
If there is an error with the controller; the error mode will start with an alarm. The
controller will revert back to Waiting Mode when the user presses START/STOP.
> Error Mode
Segment
Error 10
(Cuff
connection
error)
This error mark will be
on the screen if there
is no cuff connection
to the controller.
Lower pressure error
Under the 30
%
pressure
has consecutively
checked five times.
Vent error
Pressure in cuff doesn't
ventilate.
Pump error
When the device pump
has an error.
Device error
When the device has an
error.
The voltage of the
controller has an error
after AC power is on.
Warning of low battery
Connect the AC power.
High pressure error
Over the 30
%
pressure
has consecutively
checked five times.
Error 30
(system
error)
Error 20
(Pressure
error)
Error 40
(Power
error)
Error information
Error information
Segment
DS MAREF
11