2. Foot Pressure Button
This button sets the pressure at which the device will be operated. When the user
presses the START/STOP button, the pressure will be selected according to the type of
cuff. (Foot Cuff: 120mmHg). The pressure setting cannot be changed during operating.
Basic pressure is set as 120mmHg. The pressure setting can be changed by 10mmHg
increments by pressing the button (ex. 120
→
130
→
140
→
120mmHg). The LCD screen
will remain be off if the user does not apply the cuff , as the controller automatically
selects the cuff when the START/STOP button is selected.
The controller has an automatic gradient pressure application which applies gradient
pressure into each chamber of a cuff sequentially (+/- 5mmHg).
For cuffs with 3 chambers such as the leg/calf/boot cuff; if 40mmHg is set, pressure
will be delivered as 35/40/45mmHg to the three chambers.
3. Interval Button
This button sets the interval time. The time value which the user sets will be
saved. Default time is set at 48 sec, and it can be changed to 48
→
60
→
24
→
48 sec
each time the interval button is selected. After initial inflation of pressure to each
chamber the controller will then restart therapy with the interval time selected.
This operation will be repeated and the setting cannot be changed during operation.
DS MAREF