  Equipment Installation 

1.1

 

Installation Preparation 

 

Installation Preparation

 

 

5 meter long tape for distance measurement on site 

 

DC12V adapter, used with camera 

 

Some cable lines 

 

Product appearance 

BF3221

BF5421: 

    Blackbody

 

Note:  the  recorded  installation  guidance  video  "Dahua  indoor  human  body  temperature 

measurement  product  installation  video  (BF3221)",  which  adopts  tripod  mode,  is  applicable  to 
portable indoor installation guidance reference, and other fixed installation needs to be installed in 
strict accordance with the specified height, distance and other requirements. 

The  PPT  version  of  installation  guidance  video,  web  operation  guidance  and  guidance 

document  should  be  obtained  from  the  regional  delivery  supervisor.  In  the  future,  it  will  be 
synchronized in the group on the mobile app. 

1.2

 

Installation Distance 

focal length 

Distance between 

blackbody and 

camera 

Distance between human 

forehead and camera 

Width at the best 

distance of temperature 

measurement 

BF3221

3.5mm

 

2m 

2m 

(NO Blackbody2m) 

1.5m 

BF3221

7mm

 

3m 

3m 

(NO Blackbody2m) 

1.3m 

BF5421

13mm

 

3m 

3m 

(NO Blackbody2m) 

1.5m 

The distance from the forehead to the camera and the distance from the blackbody to the camera have the 
best  temperature  measurement  accuracy.  Taking  BF3221  (7mm)  as  an  example,  the  distance  between 

 

 

Blackbody radiant surface 

 

Summary of Contents for BF Series

Page 1: ...Dahua Thermal Body Temperature Measurement Solution Installation Guide BF Series Zhejiang Dahua Technology Co Ltd...

Page 2: ...ture exceeds 35 there is a possibility of fever For other environmental temperature conditions it is recommended to measure several samples on site and increase 1 as the alarm threshold on the basis o...

Page 3: ...Preface II Change Record Revision date Version number Creater Reviser 2020 01 27 V1 0 2020 01 28 V1 1 2020 01 31 V1 2 2020 02 03 V1 3...

Page 4: ...2 General Settings 7 2 1 Equipment Upgrade 7 2 2 Debugging Of Human Body Temperature Detection Function 8 2 2 1 Selection of human body temperature measurement intelligent scheme 8 2 2 2 Blackbody Par...

Page 5: ...nd other requirements The PPT version of installation guidance video web operation guidance and guidance document should be obtained from the regional delivery supervisor In the future it will be sync...

Page 6: ...by one at the designated position BF5421 13mm installation requirements are consistent with BF3221 7mm 1 3 Installation Diagram BF3221 3 5mm installation diagram Blac kbod y Height 2m Side view h u m...

Page 7: ...objects 3m Height of Blackbody 1 8m Thermal camera focal length 13mm Blac kbod y Sidewalk width 1 5m Top view Thermal camera focal length 13mm 1 4 Scene Diagram BF3221 7mm as an example Distance betwe...

Page 8: ...Height of Blackbody 1 8m Personnel access direction 1 When the artificial diversion channel is lengthened as far as possible multiple S shaped can be used When the indoor and outdoor temperature diffe...

Page 9: ...e facing the camera irradiation direction the black body shall be on the left or right side of the thermal imaging picture and the tripod can be protected by temporary fence As follows blackbody is in...

Page 10: ...tified in the delivery group please pay attention to the group information in time The default to be restored after upgrading is as follows Note 1 If temperature compensation has been carried out in 2...

Page 11: ...blackbody one is the digital tube display right two the other is the LCD display left Note For the blackbody display of the LED just press the bottom to adjust the value for the LCD display device pr...

Page 12: ...fter drawing the wireframe should be drawn along the edge of the bright area before 20200210 the wireframe should be in the middle of the blackbody the smaller the wireframe the higher the measurement...

Page 13: ...in the figure below indicates that an audible and visual alarm will be generated when the head temperature of the subject exceeds 37 3 When the capture angle filter value is 1 only the front side simi...

Page 14: ...e Thermal Camera Face Frame Camera settings camera properties channel 2 find binocular calibration adjustment adjust the up down left and right arrows set step length according to the actual situation...

Page 15: ...12 3 Announcements The following is for blackbody Warning A warning indicates a condition that may cause harm to the user To avoid personal injury or instrument...

Page 16: ...iation strong electromagnetic interference and vibration Heat dissipation space shall be reserved around the blackbody and the distance from the surrounding objects shall not be less than 10cm Calibra...

Page 17: ...nt is 10 30 6 The best distance for temperature measurement The best temperature measurement distance is the same as the distance between blackbody and equipment 3 5mm detection distance is 2m and the...

Page 18: ...tion alarm Select event management blackbody exception alarm blackbody exception enable and turn on or off The default sensitivity is 60 and the lower the value the less likely it is to report blackbo...

Reviews: