3. Product Usage
3.3 Usage Instructions
-
If turning ON the power of this product, please turn ON the Main Power Switch after turning
ON the Leakage Breaker located at the bottom of the rear panel.
-
If turning OFF the power of this product, please turn OFF the Leakage Breaker after turning
OFF the Main Power Switch. If you turn OFF the Leakage Breaker without turning OFF the
Main Power Switch, the operation of the controller will be continued by the battery built into
this product. This system is prepared for the controller being operated by the battery
built into this product, and the alarm and LCO
2
spraying can be operated when there is a
problem in the power supply such as a power outage.
-
Please start operation after at least one minute when you turn ON the power supply again.
①
Set temperature
-
Following to supply electricity power to main controller, the freezer goes into
“Operation Mode” with display of present temperature and set temperature.
-
Push “Set” button, then SV display is blinking and ready to set.
-
Push again “Set” button to change position of set value and Push “
▲
” and “
▼
”
button to increase and decrease the value. (the temperature is already set when
it is released in factory)
-
After completing to input the set value, press “Set” button to go back to
“Operation Mode”. The freezer will have stabilization time for 1 minute and will
start to operate.
* Caution
1) The freezer may have stabilization time for 1 minute in operation mode. The delay
time is designed and required by the reason of compressor safety. If it takes over a few
minutes, please contact the nearest our official distributor or DAIHAN Scientific Co. Ltd.
directly.