<Fig.3.3.15> Details in the Graph
<Fig.3.3.16> Setting of Offset Values (for Calibration)
-
By using the rightward or leftward arrow buttons in the display of details, the cursor can
make fine move along the points on the graph. The part of the display of details is the
vertical line of orange color that is indicated as a straight line on the whole graph.
3.3.4. OFFSET
-
When the two thermometers that are displayed in the controller of this device and the one
prepared for an experiment are to be compared and corrected thereby, the value needed
for the correction thereof can be set.
-
The value of correction ranges from -15.0
℃
to +15.0
℃
, and the current temperature to
be displayed after the correction will reflect the correction made therefor.