DAIHAN Scientific, Ultra Low Temperature Smart Freezer
36/79
<Fig.3.3.21> Import Completed
<Fig.3.3.22> If the folder or file name
incorrect, Not recognized
<Fig.3.3.23> Overwrite confirmation message
-
On completion of the “Import”, the ‘Storage’ information in the device will be deleted and
then be replaced with the information contained in the imported Excel file.
3.3.6. LCO
2
Injection
-
When the temperature of the instrument is suddenly increased due to power outage, etc.,
damage to valuable samples can be prevented by LCO
2
.
-
This operation is activated when reaching the setting temperature, and if the temperature
becomes higher than the setting temperature, LCO
2
is automatically sprayed into the
instrument by “Temp Error” setting, and LCO
2
is sprayed while the LCO
2
button on the
screen is pressed, if necessary.
-
If Freezing System error occurs in situations that the user cannot control, the light is turned
on automatically, and LCO
2
is sprayed until the error is solved.
-
When you open the door during operation, this function becomes disabled, so LCO2
spraying with the door open is prevented. When the door is closed and the objective
temperature is recovered, the reactivation is achieved.