DAIHAN-Sci.com or
ALL
for
LAB.com
Shopping Mall
DAIHAN Scientific, Ultra Low Temperature Smart Freezer
38/79
3.3.7. Error Message
-
If happen a error, Error notice will be showed in the display and if you click it, you can see
what happened in detail.
* When malfunctions in the Touch Screen occur, please press the Reset Button by inserting an
appropriate tool (such as a paper clip) into the Reset hole, which is located in the right frame of
the Touch Screen as shown in the figure below. The Touch Screen is rerun by pressing the
Reset Button.
ERROR
Message
Reasons
Actions
Temp Sensor
Error
This will be generated by the error of
temperature sensor.
Please refer to the A/S Center
to correct error of sensors.
Door Open
This will be generated when the door is
remaining opened more than 1 minute.
Check the status of the door.
Out of Control
– High
This will be generated by the temperature
deviation exceeding the ‘Upper Limit’ set in
the menu of ‘Temp Failure Condition’ by
users.
Check the set temperature and
set ‘Upper Limit’.
Out of Control
– Low
This will be generated by the temperature
deviation exceeding the Lower Limit set in
the menu of Temp Failure Condition by
users.
Check the set temperature and
set ‘Lower Limit’.
Compressor
#1 Error
This will be generated by the anomaly of
Compressor #1.
Please refer to the A/S Center
to correct failures of internal
components.
Compressor
#2 Error
This will be generated by the anomaly of
Compressor #2.
Please refer to the A/S Center
to correct failures of internal
components.
Power Error
This will be generated by the anomaly of
power supply.
Check the ‘Circuit Breaker’.
Control
Response
Error
This will be generated by the failure of
communication between device and ‘Touch
Board’.
Check the connection to ‘Touch
Board’.
<Table.3.3.1> Error Message Table