(2) Alarm reset
In case an alarm return method is set to
“
HOLD (manual return)
”
, an alarm output is reset (output OFF).
(With an alarm-output option)
However, an output is not turned off by this operation, in case an alarm continues and it has caused.
And, this operation is unnecessary if an alarm return method is set as "AUTO (automatic return)".
(An output is also OFF to an alarm return.)
In addition, setting can perform same operation in external operation input.
Please refer to "5.3.2 Setting mode 2 (4) external operation input setting" for the setting method.
Please refer to "6.3 Option" for the external operation input.
① It continues pushing RESET/SHIFT 1 second or more by a measurement display, the maximum measurement
display, and the minimum measurement display.
<Caution> If a RESET/SHIFT switch performs alarm reset, the maximum value of the measurement factor
currently displayed on the main monitor and the minimum value are also reset.
Reset completion
(3) Maximum value and minimum value reset.
Reset of the various measurement values of maximum value and minimum value is performed.
This reset has two types of methods. (How to perform according to a measurement factor individual.
How to reset all maximum values and minimum values by package.)
a) Individual reset
It resets the differential maximum value and minimum value. Other maximum values and minimum values
are not reset by this operation.
① It displays a measurement factor to reset on the main monitor. (By measurement display and the
maximum and the minimum measurement display.)
② Press RESET/SHIFT for longer than 1 seconds.
<Caution> Please be sure to perform this operation after displaying the maximum value and a minimum
value measurement factor to reset. And, if the maximum value and minimum value reset are
performed, the alarm output of detection will also be reset.
Maximum, Minimum measurement display Reset completion
RESET/SHIFT
1 second
Instant display
It displays in bar graph, the
range of minimum to maximum.
Minimum display Maximum display
Minimum display Instant display Maximum display
(Reset) (Reset)
RESET/SHIFT
1 second
A(V) display
V(UV) display W display Wh display
Alarm detection (Hold)
Instant measurement display