22
When the symbol
fl
ashes on screen,
press SET button to enter the Quality
Control Mode.
The screen will display the symbol .
4
3
NOTE
To avoid contamination,
do not apply the control
solution to the test strip
directly from the bottle.
2
A01
A01
CODE
Quality Control of the System
If the code number matches that on the
test strip vial, you can go to next step.
If not, refer to section "Code Setting" to
set the code number.
Shake the glucose control solution well
before squeezing.
Discard the first droplet and squeeze
the second one to your clean
fi
ngertip or
a clean object such as a transparency.
Move the test strip close to the object
(with control solution on it). The solution
droplet is automatically drawn into the
collection zone of the test strip.
The meter will then beep and begins to
count down for 5 seconds.
Insert the test strip into the meter.
1