23
Check if the value falls within the range
printed on the test strip vial.
Quality Control of the System
Understanding the Reading!
If the test reading falls within the expected range printed on the test
strip vial, then the system is working correctly. The results of your
blood glucose tests are reliable. If the test reading falls outside the
expected range, please check the following:
Are the control solutions or test strips within the expiration date?
Has the control solution been contaminated?
Is the code number correct?
Did you follow the test procedures correctly?
Please repeat the test. If the reading still is not acceptable,
c o n t a c t with your local H E A LT H & L I F E office o r t h e
authorized distributor.
5
NOTE
Low and high control range values are
printed on the test strip vial label. You may
also refer to DailyChek
®
Glucose Control
Solution instruction.
mmol/L
Summary of Contents for HL568AA
Page 1: ...User s Manual Premium Self Monitoring Blood Glucose System Premium Blood Glucose Meter...