Dakota Ultrasonics
6
F. Backlight :
When this icon is illuminated, it indicates the backlight is on.
G. Small 7 Segment:
The material velocity, speed the sound wave travels
through a given medium/material, is displayed in this area, informing the user
what material the
PZX-7
is currently calibrated too. This area is also used for
alpha messages in the menu and edit modes.
H. Units:
This combination of icons are illuminated in different sequences to
inform the user what measurement units are currently being displayed in the
small 7 segment area.
I. Small 14 Segment:
Displays the current gain setting of the
PZX-7 DL
. In P-E
mode the options are
VLOW
, LOW, MED, HIGH,
VHI
, and LOW, MED, HIGH
with modes E-E, I-E, and PLAS. MED is the default setting.
J. Units:
This combination of icons are illuminated in different sequences to
inform the user what measurement units are currently being displayed in the
large 7 segment area. The plus/minus icon is illuminated when the DIFF
(differential) feature is activated.
K. Small 14 Segment:
The material type is displayed in this area. If it is set to a
value of one of the materials in our material list, it will be displayed in alpha
characters indicating the material type. Otherwise it will be set to CUST,
indicating custom material type.
L. Features:
The icons illuminated in this row across the bottom of the LCD
display which features are currently enabled. For a complete list of the menu
features in the
PZX-7
, Refer to page 4 for a list. The
PZX-7
can be locked
once calibrated, to avoid accidently changing the calibration. When this icon is
illuminated, the
PZX-7
is in lock mode. Refer to page 38 for an explanation on
locking the
PZX-7
.
Summary of Contents for PZX-7