On Block Zero (Manual Probe Zero)
– When this feature is enabled the
transducer must be placed on the probe zero disk (battery cover) located on
the top of the unit.
Both zero procedures are outlined as follows:
Performing an Auto Probe Zero (Off Block)
1) Press the
key to perform the auto zero.
“AUTO”
will be displayed on
the screen and flashing
CLn/Prb
(clean probe).
2) Make sure the couplant is wiped clean from the tip of the transducer to avoid
any zero error.
3) Press the
key to perform the zero.
Performing a Manual Probe Zero (On Block)
Note:
When the zero probe option is set to manual, the probe zero disk
(battery cap) located on the top of the gauge will be used as a zero standard.
1) Apply a drop of couplant on the transducer and place the transducer in
steady contact with the disk (battery cover) located at the top of the unit to
obtain a measurement.