Dakota Ultrasonics
24
1) Press the
key to edit the gain setting. The edit icon will be illuminated
and flashing.
2) Use the
keys to scroll through the gain settings
(VLOW,
LOW, MED, HIGH, VHI)
until the desire setting is being displayed.
3) Press the
key to set the gain and return to the measurement screen.
Note:
Pressing the
key prior to pressing the
key will abort to the
measurement screen without saving changes.
6.2 High Speed Scan
The High Speed Scan feature of the
ZX-3
increases the overall repetition rate to a
maximum of 140Hz with a high speed screen refresh rate of 25 times a second. This
allows for making scanned passes over an arbitrary length of the test material, while
still maintaining a reasonable representation of thickness over the area or region
scanned.
The procedure to use the scan feature is outlined below:
High Speed Scan
1) Press the
key to toggle
SCAN
on/off. The display will briefly display the
status and return to the measurement screen.
6.3 Units
The
ZX-3
will operate in both English (inches) or Metric (millimeters) units.
www.calcert.com
sales@calcert.com
1.888.610.7664
0
5
10
15
20
25
30