29
Program values
Calibration check
Use the program Values to check if the detector readings are within
specified tolerances.
Quick check
1. Set the tolerance to 0.01 mm (0.5mil).
2. Select
and show targets for both M- and S-unit.
3. Select
to zero set value.
4. Place a shim under the magnet base to lift the M-
unit 1mm (100mils). The M-unit’s reading shall
correspond to the movement within 1% (1mil ±
1digit) (0.01mm ± 1 digit).
5. Remove the shim from the M-unit.
6. Select
to zero set value.
7. Make a mark to mark out the position of the detec-
tor.
8. Place the shim under the magnet base of the S-unit.
The S-unit’s reading shall correspond to the move-
ment within 1% (1mil ± 1digit) (0.01mm ± 1 digit).
Note!
The shim must be exactly 1 mm. In this example it is only the M-unit that is checked.
Select to show both targets.
Zero set value
Precision check
1. Fasten one unit in a machine tool.
2. Select
to zero set value.
3. Move the units a known distance is to use the move-
ment of a machine tool spindle.
4. The fastened unit’s reading shall correspond to the
movement within 1% (1mil ± 1digit) (0.01mm ± 1 digit).
Note!
In this example it is only the unit fastened in the machine that is checked.
Parallel lift to a known distance.
Shim exactly 1mm.
Make sure that the distance is kept
www.
GlobalTestSupply
.com
Find Quality Products Online at:
sales@GlobalTestSupply.com
Summary of Contents for Easy-Laser E420
Page 2: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...
Page 26: ...22 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...
Page 28: ...24 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...
Page 34: ...30 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...
Page 58: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...