User Guide
MES 1003 WMS
Doc no. 100837-MAN – Rev. E1 – Rev. date 2022-10
© Danfoss IXA A/S | Page 51
Verification of PAH measurement
1.
Activate verification mode as described in 5.4.1
2.
Remove the bottom half of the measuring chamber as described in 5.2.2
3.
Clean the optical part of the PAH sensor with clean water and a lint free cloth
4.
Place the DryCal-0 or DryCal-1 on the PAH sensor as shown on in the illustration below
5.
Verify the measurement on the LCP
6.
Reinstall the bottom half of the measuring chamber
7.
Deactivate verification mode as described in 5.4.1
Illustration 88: Placing the DryCal-0 or DryCal-1 on the PAH sensor for verification