User Guide
MES 1003 WMS
Doc no. 100837-MAN – Rev. E1 – Rev. date 2022-10
© Danfoss IXA A/S | Page 40
5.3.2.
Turbidity Sensor Calibration
The turbidity sensor can be re-calibrated with a one-point calibration for scaling. The calibration must be performed
following the procedure as described in this section.
Tools needed for turbidity calibration:
Allen key size 5
TTurbCAL
Lint free cloth & clean water
Incorrect calibration always leads to incorrect measured values. Always use a TTurbCAL paired with the specific turbidity
sensor when calibrating to prevent an incorrect calibration.
Switch off the power for the WSU pumps, close all isolation valves for the WMS system and drain the measurement
chamber before starting the turbidity calibration.
Turbidity calibration
To start the turbidity calibration, go to the Turb. Calibration menu 6-3*. The following illustrations shows LCP when
performing the turbidity calibration.
Illustration 52: Maintenance submenu
Press OK to enter the Turb. Calibration menu.
Illustration 53: Turb. Calibration menu
Press OK to activate the Turb. calibration mode.
Maintenance
6-**
6-1*
Verification Mode
6-2*
pH Calibration
6-3*
Turb. Calibration
6-3*
PAH Calibration
Turb. Calibration
6-**
Turb. Calibration
Press OK to activate
calibration mode